Safety and efficacy of a new device for inducing ejaculation in men with spinal cord injuries
- Case Report
- Open access
- Published:
- S M Castle1,
- L C Jenkins1,
- E Ibrahim1,
- T C Aballa1,
- C M Lynne1 &
- …
- N L Brackett1
Spinal Cord volume52,pages S27–S29 (2014)Cite this article
-
52k Accesses
-
31 Citations
-
2 Altmetric
-
Metrics details
Subjects
- Clinical trials
- Infertility
- Spinal cord diseases
Abstract
Study design:
Prospective case series.
Objectives:
Male infertility is a common sequela of spinal cord injury (SCI). Fatherhood is a goal in this group of young patients; however, most are anejaculatory. Penile vibratory stimulation is recommended as the first line of treatment for this condition. Our study evaluated the safety and efficacy of a new device designed to induce ejaculation in these patients.
Setting:
The Miami Project to Cure Paralysis, Miami, FL, USA.
Methods:
The Viberect-X3 (Reflexonic, Frederick, MD, USA) was applied to 30 consecutive anejaculatory men with SCI whose level of injury was T10 and rostral.
Results:
The ejaculatory success was 77% (23/30). No adverse events occurred, and there were no malfunctions of the device.
Conclusion:
In this first report on the efficacy of the Viberect-X3 for treatment of anejaculation in men with SCI, we conclude that the device is safe and effective for inducing ejaculation in men with SCI. Recommendation of the Viberect-X3 versus other devices intended for this purpose should not be made until randomized controlled trials are performed.
Similar content being viewed by others
Urethral pressure profile during ejaculation in men with spinal cord injury
Article 13 January 2022
Evaluation of a re-engineered device for penile vibratory stimulation in men with spinal cord injury
Article 14 July 2020
Use of the ice water test as an early predictor of recovery of erectile function in patients with spinal cord injury
Article 29 June 2020
Introduction
Every year, there are ∼12 000 new cases of spinal cord injury (SCI) in the United States,1 and thousands more worldwide. Most spinal cord injuries occur in young males.1 Following SCI, the majority of men become anejaculatory and require medical assistance to procure sperm.2 Only 10% of men with SCI can ejaculate via masturbation.3
Owing to its safety, cost and reliability, penile vibratory stimulation (PVS) is recommended as the first line of therapy for anejaculation in men with SCI.3 PVS typically is performed by applying a single vibrator to the dorsum or frenulum of the glans penis. During any PVS procedure, care must be taken to manage adverse symptoms such as autonomic dysreflexia and penile skin edema/abrasion.4, 5 Historically, a wide variety of ejaculation success rates (19–96%) have been reported from PVS of men with SCI, with the highest success rates achieved by devices capable of delivering amplitudes ⩾2.5 mm and 100 Hz, applied to men whose level of injury was T10 and rostral.3, 6, 7
Failures (22%) may be salvaged by ‘sandwiching’ the penis between two vibrators;8 however, this maneuver can be cumbersome to perform. A new vibrator has been developed to achieve, in a single device, simultaneous stimulation of the dorsum and frenulum of the penis (Viberect-X3, Reflexonic, Frederick, MD, USA; Figure 1). To date, there are no reports on the efficacy of this device for inducing ejaculation in men with SCI. The purpose of our study was to evaluate the ejaculatory success rate and safety of the Viberect-X3 in anejaculatory patients with SCI.
Materials and methods
The study was approved by our Institutional Review Board. Subjects were men with SCI enrolled in the Male Fertility Research Program of the Miami Project to Cure Paralysis. All signed informed consent documents. We prospectively studied 30 consecutive SCI subjects who were unable to ejaculate by sexual intercourse or masturbation. Ejaculatory response to PVS relies on an intact spinal reflex arc;3 therefore, subjects with injuries caudal to T10 were excluded from the study. All patients received one trial of PVS with Viberect-X3. All patients were familiar with PVS and had been administered one or more previous trials with an alternate device. The present study was not designed to compare the success rates of various PVS devices, but simply to assess the ejaculation success rate with Viberect-X3.
Prior to PVS, subjects whose level of injury was T6 or rostral were administered 10–40 mg nifedipine sublingually to manage autonomic dysreflexia. Viberect-X3 was administered at the non-adjustable manufacturer settings of 4 mm amplitude and 70–100 Hz. Based on our previous studies, PVS was performed in 2–5 min intervals, interspersed by 1–2 min rest periods, for up to three intervals of stimulation, or until ejaculation occurred.3 Blood pressure was recorded at 1 min intervals throughout the procedure. Each patient was monitored for adverse symptoms, including penile bleeding or edema. The vibrator pads were thoroughly cleansed with 70% alcohol between each patient use.
Results
Details of patient information are listed in Table 1. The result of the PVS procedures was that 23 of the 30 patients (76.7%) ejaculated. No adverse symptoms were noted. All patients tolerated the procedure well. The Viberect-X3 device did not malfunction during any of the stimulations.
Discussion
PVS is thought to cause ejaculation via stimulation of the dorsal penile nerve.9 In order to achieve ejaculation, the penile vibrator must appropriately stimulate this nerve. Earlier reports showed low-amplitude vibrators to be inferior to high-amplitude vibrators for inducing ejaculation in men with SCI.10 Viberect-X3 deploys high-amplitude (4 mm) vibration via two sites in a single device. Viberect-X3 represents a new option for these patients. The success rate obtained in our study (76.7%) was slightly lower than previously published PVS success rates.3 This new penile vibrator appears to be safe and reliable in our cohort. For practitioners who have administered PVS with other devices, the learning curve is quite short, that is, one to two trials should be sufficient to successfully administer Viberect-X3. Prior to the patient self-administering this device at home, an experienced practitioner should supervise the patient’s ability to safely and effectively administer the Viberect-X3.
Previous research has described maneuvers to salvage patients who fail to ejaculate with the administration of one single-headed vibrator. Such maneuvers include sandwiching the penis between two single-headed vibrators, administration of abdominal electrical stimulation in conjunction with PVS and administration of medications such as midodrine (for a review, see Brackett et al.3). The current study did not assess the ability of Viberect-X3 to salvage failures to a single-headed vibrator.
In this first report on the efficacy of Viberect-X3 for treatment of anejaculation in men with SCI, we conclude that the device is safe and effective for inducing ejaculation in men with SCI at or rostral to T10. Recommendation of Viberect-X3 versus other devices intended for this purpose should not be made until randomized controlled trials are performed.
Data archiving
There were no data to deposit.
References
National SCI Statistical Center Spinal Cord Injury—Facts and Figures at a Glance, 2013 available from https://www.nscisc.uab.edu/PublicDocuments/fact_figures_docs/Facts%202013.pdf.
Brown DJ, Hill ST, Baker HW . Male fertility and sexual function after spinal cord injury. Prog Brain Res 2006; 152: 427–439.
Brackett NL, Lynne CM, Ibrahim E, Ohl DA, Sonksen J . Treatment of infertility in men with spinal cord injury. Nat Rev Urol 2010; 7: 162–172.
Elliott S, Krassioukov A . Malignant autonomic dysreflexia in spinal cord injured men. Spinal Cord. 2006; 44: 386–392.
Brackett NL, Lynne CM, Sonksen J, Ohl D . Penile vibratory stimulation. In: Goldstein M, Schlegel PN (ed). Surgical and Medical Management of Male Infertility 1st edn. Cambridge University Press: New York. 2013 pp 119–129.
Sonksen J, Ohl DA . Penile vibratory stimulation and electroejaculation in the treatment of ejaculatory dysfunction. Int J Androl 2002; 25: 324–332.
Sonksen J, Biering-Sorensen F, Kristensen JK . Ejaculation induced by penile vibratory stimulation in men with spinal cord injuries. The importance of the vibratory amplitude. Paraplegia. 1994; 32: 651–660.
Brackett NL, Kafetsoulis A, Ibrahim E, Aballa TC, Lynne CM . Application of 2 vibrators salvages ejaculatory failures to 1 vibrator during penile vibratory stimulation in men with spinal cord injuries. J Urol 2007; 177: 660–663.
Wieder J, Brackett N, Lynne C, Green J, Aballa T . Anesthetic block of the dorsal penile nerve inhibits vibratory-induced ejaculation in men with spinal cord injuries. Urology 2000; 55: 915–917.
Brackett NL, Ferrell SM, Aballa TC, Amador MJ, Padron OF, Sonksen J et al. An analysis of 653 trials of penile vibratory stimulation in men with spinal cord injury. J Urol 1998; 159: 1931–1934.
Author information
Authors and Affiliations
The Miami Project to Cure Paralysis, University of Miami Miller School of Medicine, Miami, FL, USA
S M Castle,L C Jenkins,E Ibrahim,T C Aballa,C M Lynne&N L Brackett
Authors
- S M Castle
View author publications
You can also search for this author in PubMedGoogle Scholar
- L C Jenkins
View author publications
You can also search for this author in PubMedGoogle Scholar
- E Ibrahim
View author publications
You can also search for this author in PubMedGoogle Scholar
- T C Aballa
View author publications
You can also search for this author in PubMedGoogle Scholar
- C M Lynne
View author publications
You can also search for this author in PubMedGoogle Scholar
- N L Brackett
View author publications
You can also search for this author in PubMedGoogle Scholar
Corresponding author
Correspondence to S M Castle.
Ethics declarations
Competing interests
The authors declare no conflict of interest.
Rights and permissions
This work is licensed under a Creative Commons Attribution 3.0 Unported License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/3.0/
About this article
Cite this article
Castle, S., Jenkins, L., Ibrahim, E. et al. Safety and efficacy of a new device for inducing ejaculation in men with spinal cord injuries. Spinal Cord 52 (Suppl 2), S27–S29 (2014). https://doi.org/10.1038/sc.2014.110
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/sc.2014.110
This article is cited by
-
The sextech industry and innovative devices for treating sexual dysfunction
- Shelly Varod
- Rafi J. Heruti
International Journal of Impotence Research (2023)
-
A rare cause of anejaculation: mitochondrial neurogastrointestinal encephalomyopathy (MNGIE) syndrome: case report
- Selman Unal
International Journal of Impotence Research (2023)
-
Treatment of Delayed Ejaculation
- Kristen Piche
- Uday Mann
- Premal Patel
Current Sexual Health Reports (2020)
-
Orgasm and SCI: what do we know?
- Marcalee Alexander
- Lesley Marson
Spinal Cord (2018)
-
Comparison of three methods of penile vibratory stimulation for semen retrieval in men with spinal cord injury
- W Chong
- E Ibrahim
- N L Brackett
Spinal Cord (2017)
Associated content
Special
August 2014, Volume 52 No S2
Supplement
Series and Cases Supplements Collection
Advertisement